Dual-Mode Ultrasound Array Demonstrates Focused Electronic Steering for
A preclinical sixty-four-element ultrasound array transducer developed for targeted drug delivery in abdominal organs successfully demonstrated functional electronic focus steering within a steerable volume of approximately one cubic centimeter, according to e
Illustrative imageA preclinical sixty-four-element ultrasound array transducer developed for targeted drug delivery in abdominal organs successfully demonstrated functional electronic focus steering within a steerable volume of approximately one cubic centimeter, according to experimental characterizations and numerical simulations published in Acoustical Physics. The diagnostic-type linear phased array, operating at a frequency of 1.071 megahertz, was engineered to combine high-pressure therapeutic output capable of inducing localized cavitation with acoustic imaging capabilities for treatment visualization. While the physical transducer matched computer model predictions along its central acoustic axis, experimental acoustic holography revealed that its transverse steering range was reduced by half relative to an idealized numerical design. Researchers established that this limitation stemmed from reduced individual element directivity, dissipation across the focusing acoustic lens, and inter-element cross talk, while validating that synthetic summation of single-element acoustic holograms accurately reproduces complex multi-element acoustic fields.
Multi-element phased arrays are widely utilized across modern diagnostic imaging and therapeutic ultrasound procedures, yet standard clinical systems typically rely on distinct and physically separated transducers due to divergent geometric and acoustic requirements. Standard diagnostic abdominal probes operate at higher frequencies, generally between four and seven megahertz, featuring smaller apertures configured to optimize spatial resolution rather than acoustic intensity. In contrast, therapeutic transducers engineered for sonicating deep abdominal targets, such as pancreatic tumors, typically function at lower frequencies between one and one and a half megahertz, possessing larger apertures designed to deliver substantial acoustic energy to tight focal volumes. Conventional image-guided therapy setups position a compact diagnostic probe within a central opening cut into a larger therapeutic array. The current investigation evaluated an integrated linear design intended to generate sufficient acoustic pressures to induce de novo cavitation without requiring microbubble contrast agents, while simultaneously providing sonoporation monitoring at 1.071 megahertz.
To evaluate the baseline acoustic behavior and determine the optimal front-face geometry, computational investigators initially designed two distinct numerical configurations: a completely planar sixty-four-element array and a cylindrically curved alternative. The simulated aperture dimensions spanned 14.8 millimeters in elevation and 51.2 millimeters along the azimuth, consisting of sixty-four rectangular piezoelectric strips that measured 14.8 millimeters by 0.72 millimeters with an inter-element kerf of 0.08 millimeters. Numerical modeling carried out in simulated water at twenty degrees Celsius demonstrated clear acoustic advantages for the curved geometry. Cylindrical focusing along the elevation coordinate enhanced the focal peak pressure gain by 17.6 percent compared with the planar design, simultaneously depressing undesirable near-field pressures. Furthermore, evaluating the focal dimensions at the negative six-decibel threshold revealed that the cylindrical curvature contracted the transverse focal dimensions by four percent along the azimuthal x-axis and by forty-five percent along the elevational y-axis, yielding a tighter sonication zone favorable for localized therapeutic action.
Following these preliminary computational assessments, a physical prototype array was manufactured by Sonic Concepts according to the design-stage specifications, employing a cylindrical acoustic lens bonded over planar piezoelectric elements to circumvent the manufacturing complexities associated with fabricating curved ceramic elements. To investigate the true oscillatory behavior of this manufactured device, acoustic holography scans were recorded in an acoustic water tank with ambient density of 997.6 kilograms per cubic meter and small-signal sound speed of 1482.4 meters per second. A hydrophone mapped the complex pressure amplitude across a scanning plane located thirty millimeters away from the transducer face. Spatial scanning windows measured plus or minus 58.5 millimeters along the x-axis and plus or minus 22.75 millimeters along the y-axis when driving the full array, while single-element scans spanned plus or minus 55.25 millimeters by plus or minus 19.5 millimeters, capturing the complete acoustic wavefield across fine calculation steps.
Acoustic holograms recorded under full-aperture operation were pre-processed using the Rayleigh integral to compensate for experimental misalignment, applying rotation corrections of negative two degrees about the x-axis, 0.06 degrees about the y-axis, and 0.6 degrees about the acoustic z-axis. The angular spectrum method was subsequently applied to backpropagate acoustic fields from the recording plane to the active radiating face at z equals zero millimeters. When driven in an in-phase configuration without electronic steering, the reconstructed surface boundary conditions exhibited uniform pressure amplitudes, flat phase profiles along the azimuthal coordinate, and symmetric elevational phase curvature imparted by the lens. However, when electronic delays were applied to focus the beam axially at fifty millimeters or to steer it obliquely to an azimuthal coordinate of ten millimeters, noticeable amplitude non-uniformity emerged across peripheral elements, exhibiting up to twofold amplitude reductions that investigators attributed to longer, oblique propagation paths through the dissipative lens material.
To resolve the precise physical mechanisms responsible for unexpected field asymmetries observed in forward acoustic simulations, the researchers recorded sixty-four separate acoustic holograms, isolating each individual piezoelectric element while keeping adjacent channels inactive. Detailed processing of these individual records quantified performance variability across the array. The mean acoustic power emitted per element under standardized measurement conditions was 111 milliwatts with a standard deviation of nine milliwatts. Measuring spatial beam widths at a distance of fifty millimeters from the transducer surface yielded an average azimuthal width of 44.4 millimeters with a standard deviation of 2.4 millimeters along the x-axis, alongside an elevational width of 5.23 millimeters with a standard deviation of 0.01 millimeters along the y-axis. These variations in element directivity and power output were identified as primary contributors to the pre-focal beam distortion detected during steered operations.
A granular inspection of surface pressure amplitude distributions on individual elements revealed subtle acoustic discrepancies when contrasted with the uniformly vibrating pistons assumed in idealized numerical models. Elements exhibiting minimum output, average power, and maximum power all displayed slightly expanded radiating surface areas and minor amplitude ripples across their faces. When examining the directivity patterns of individual elements in the azimuthal xz-plane at a propagation distance of fifty millimeters, the generated acoustic beam width was approximately twofold narrower than the directivity pattern predicted for an idealized uniformly vibrating element, directly constraining the wide-angle acoustic emission necessary for off-axis focus steering.
The study demonstrated that synthesizing the complete array sound field by numerically superimposing the reconstructed individual element holograms with appropriate phase delays achieved remarkable accuracy. When phase corrections corresponding to an on-axis focal distance of fifty millimeters were applied to the sixty-four single-element holograms, the resulting synthesized surface boundary conditions and forward-propagated fields matched the direct full-aperture experimental measurements in both the azimuthal and elevational planes. Forward propagation calculations executed through the Rayleigh integral confirmed that inter-element electrical and mechanical coupling effects were fully preserved within the individual holography measurements. This synthetic methodology eliminated the requirement to record physical holograms for dozens of individual focal targets, allowing comprehensive computational sweeps across three-dimensional steering space based entirely on sixty-four baseline single-element measurements.
Utilizing this validated synthetic reconstruction technique, the researchers systematically mapped the electronic focus steering range across a spatial calculation grid with 0.1-millimeter step sizes, defining the operational steering boundary as the spatial contour where peak focal pressure amplitude remained within ten percent of the nominal on-axis focal pressure at fifty millimeters. Along the acoustic z-axis, the manufactured transducer matched the ideal theoretical model almost exactly, maintaining effective focusing across axial distances from 45 to 55 millimeters. However, in the transverse focal plane at z equals fifty millimeters, the physical array achieved an operational steering range of twenty millimeters (spanning plus or minus ten millimeters from the central axis), whereas the idealized computer model predicted a steering range of forty-four millimeters (spanning plus or minus twenty-two millimeters). This twofold reduction in transverse steering range directly mirrored the narrowed directivity profiles measured for individual piezoelectric strips.
Despite the narrowed transverse steering envelope, two-dimensional focal peak pressure maps calculated across the azimuthal imaging plane confirmed that the manufactured transducer generates a continuous steerable therapeutic volume of one cubic centimeter. The researchers concluded that while acoustic lens attenuation and element cross talk constrain wide-angle electronic deflection, the fabrication of dual-mode linear arrays operating near one megahertz remains technically feasible, providing a viable path toward combined high-power therapeutic sonication and real-time cavitation visualization using single-element acoustic holography for comprehensive transducer calibration.
Subject of Research: Electronic focus steering evaluation of a dual-mode therapeutic and diagnostic linear ultrasound array
Article Title: Electronic Focus Steering Capabilities of a Diagnostic-Type Linear Ultrasound Array Designed for High Power Therapy and Its Visualization
Article References: Nartov, F. A., Williams, R. P., & Khokhlova, V. A. (2024). Electronic Focus Steering Capabilities of a Diagnostic-Type Linear Ultrasound Array Designed for High Power Therapy and Its Visualization. Acoustical Physics, 70(1), 165-174. https://doi.org/10.1134/s1063771023601292
Image Credits: AI Generated
DOI: 10.1134/S1063771023601292
Keywords: ultrasound phased array, acoustic holography, electronic focus steering, drug delivery, acoustic cavitation, Rayleigh integral, angular spectrum method, transducer characterization, piezoelectric elements, acoustic lens, focal pressure, directivity pattern, pancreatic tumor therapy, element cross talk, elevational focusing, spatial beam width
News Source: Louis Brooks. (November 24, 2026). Dual-Mode Ultrasound Array Demonstrates Focused Electronic Steering for Abdominal Drug Delivery. Scienmag.